51˶

Study: More Than 1 in 3 Heater-Cooler Devices Contaminated

— For 15% of samples, bacteria and fungi levels were off the charts

MedpageToday

This article is a collaboration between 51˶ and:

Testing of heater-cooler units used in open heart surgery often turned up Mycobacterium chimaera -- an organism linked to fatal patient infections -- as well other bacteria and fungi, despite decontamination attempts.

Among samples sent to one specialty testing laboratory from 89 heater-cooler devices at 23 centers, 51% tested positive for nontuberculous mycobacteria and M. chimaera.

Four units were also colonized with Legionella, John Rihs, vice president of laboratory services at Special Pathogens Laboratory in Pittsburgh, reported at the meeting in Portland.

Of the 653 samples cultured from July 2015 through December 2016, 15% were so contaminated with bacteria and fungi, with heterotrophic plate counts up to five million CFU/mL, that initial results were uninterpretable.

The other species recovered from these units, such as M. abscessus/chelonae and M. gordonae, have not been associated with disease in this setting, Rihs said in an interview (which was monitored by conference media relations).

"But if it's raining down M. chimaera over the surgical field, it's likely raining down those too," Rihs told 51˶, noting that such infections have probably occurred without being connected to the devices.

Nontuberculous mycobacteria are ubiquitous in the environment and particularly endemic in plumbing but typically grow slowly, such that patients may not present with problems for years after the surgery during which infection occurred.

The problem with M. chimaera in heater-cooler units emerged in 2015 with case reports from Europe, subsequently found in many cases to be tied to the popular LiveNova/Sorin 3T devices that had been contaminated at the manufacturing plant during testing and shipped wet to hospitals around the world.

However, all commercially-available heater-cooler devices have had reports of contamination sent to the FDA through mandatory or voluntary reporting.

Once colonized and a biofilm forms, it's difficult to eradicate, Rihs noted.

Eleven of the units originally clear of M. chimaera became positive later in his study, and some decontaminated devices then tested positive again months later.

The FDA has warned against using contaminated 3T devices and suggested moving away from those manufactured before the source contamination was eradicated. But it didn't recommend testing all machines for M. chimaera because false negative rates are high.

Relatively few labs can do the water sample testing required, Rihs said. Still, given the high contamination rates, he called the FDA stance on routine testing "a little disturbing."

"The only way you're going to know these units are not colonized is to do routine testing," he said. And even with updated instructions for use and cleaning to help reduce risk, "without monitoring these microbiologically, you'll never know if a break in [decontamination] technique has occurred."

Disclosures

Rihs disclosed no relevant relationships with industry.